Product Overview
The Automatic Cleveland Open Cup Flash Point Apparatus are suitable for the determination of petroleum products according to the standard petroleum products flash point and ignition point determination method (Cleveland open cup method) except for fuel oil and petroleum products with open flash point lower than 79℃.
The instrument is designed as a table-top structure, with stainless steel working table and stainless steel heating furnace shell, which is beautiful and elegant. The Automatic Cleveland Open Cup Flash Point Apparatus are equipped with flame extinguishing cover, which meets the requirements of the ignition point test.
Features
We believe that our Automatic Cleveland Open Cup Flash Point Apparatus can compete with any other similar products in terms of quality performance as well as price. It adopts a special heating furnace form to ensure the safety of the test. And its heating power is continuously adjustable to meet the requirements of the test. In addition, the instrument is complete, the user only needs to match the gas or other civil gas can be tested, it is very simple and convenient to use.
HZKS-3 Automatic COC Cleveland Open Cup Flash Point Apparatus
Technical Parameters
Temperature Measurement | |
Temperature measuring range |
-59.9°C~399.9°C |
Temperature Control Range |
Room temperature~400°C |
Repeatability |
≤150℃ 3℃, >150℃ 5℃ |
Resolution |
0.1°C |
Precision |
0.5% |
Temperature Sensor |
Platinum Resistor(PT100) |
Flash Fire Sensor |
Ion detection ring |
Ambient Temperature |
0 ~ 40°C |
Relative Humidity |
≤ 85% RH |
Power Supply |
AC220V ± 20%; 50HZ ± 10% |
Power Consumption |
Approx 400W |
Overall Dimension |
440X280X290(mm) |
Instrument Weight |
14kg |
Failure And Treatment
NO. |
Failure |
Treatment |
1 |
Turn on the power, the LCD screen no display |
Check power supply is plugged in, open the instrument measured the board examines the plug is not loose. |
2 |
Large repeatability error for sample testing |
Remove the lifting arm top cover, and check |
3 |
No heating |
Electric wire broken |
4 |
Lift arm rise or fall failure |
Limit opt coupler damage |